6 Measurements and Math Functions
204 InfiniiVision Oscilloscopes User’s Guide
The example below shows a delay measurement between the rising edge of channel 1
and the rising edge of channel 2.
Phase
Phase is the calculated phase shift from source 1 to source 2, expressed in degrees.
Negative phase shift values indicate that the rising edge of source 1 occurred after the
rising edge of source 2.
1 Press the Settings softkey to select the source 1 and source 2 channels for the phase
measurement.
The default Phase settings measure from channel 1 to channel 2.
Phase =
Delay
Source 1 Period
-----------------------------------------
X 360
Source 1
Source 2
Delay
Period