Triggering 4
InfiniiVision Oscilloscopes User’s Guide 129
14 Press the Alignment softkey and rotate the Entry knob to select the desired alignment
of the data signal (SDATA). The on-screen diagram changes with your selection.
Standard Alignment: MSB of data for each sample is sent first, LSB is sent last. The
MSB appears on the SDATA line one bit clock after the edge of the WS transition.
Left-Justified: Data transmission (MSB first) begins at the edge of the WS transition
(without the one-bit delay that Standard format employs).
Right-Justified: Data transmission (MSB first) is right-justified to the transition of
WS.