Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 6- 25
Remote Control Interface
WORKspace Subsystem
WORKspace Subsystem
Describes WORKspace subsystem commands used for opening/closing workspace,
and so on.
:WORKspace:CATalog?
Returns catalog of your own workspace and the Public workspace.
Syntax :WORKspace:CATalog?
Query response "workspace1"[,"workspace2"[, ... "workspace"]]<newline>
Data type is SRD.
Example :WORK:CAT?
:WORKspace:CLOSe
Closes the workspace now opened. Use the *OPC? command to confirm the
completion of close operation.
Syntax :WORKspace:CLOSe
Example :WORK:CLOS
:WORKspace:OPEN
Opens the specified workspace. Use the *OPC? command to confirm the
completion of open operation.
Syntax :WORKspace:OPEN "name"
Parameter name Workspace name. Data type is SPD.
Example :WORK:OPEN "my workspace1"