6- 12 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Remote Control Interface
BENCh Subsystem
Example :COUN 100
:COUN?
[:BENCh]:COUNt:RESet
Clears the value of the Count field.
Syntax [:BENCh]:COUNt:RESet
Example :COUN:RES
[:BENCh]:LOAD[:SETup]
Loads and sets the test setup information included in the XTS or XTR data. If the
XTR data contains multiple setup information, only the first setup information is
loaded.
Syntax [:BENCh]:LOAD[:SETup] setup
Parameter setup Test setup information. Data type is Block.
Example :LOAD #32229876543210 ... 987654321098<newline>
In this example, #3222 is followed by 222-byte length binary data.
[:BENCh]:PRESet:CATalog?
Returns catalog of the preset group (My Favorite).
Syntax [:BENCh]:PRESet:CATalog?
Query response "preset1"[,"preset2"[, ... "preset"]]<newline>
Data type is SRD.
Example :PRES:CAT?
[:BENCh]:PRESet:OPEN
Opens the specified preset group (My Favorites).
Syntax [:BENCh]:PRESet:OPEN "name"