1. Temperature: 23 °C ±5 °C
2. Humidity: 30 % to 80 % RH
3. After 60 minutes warm-up
4. Ambient temperature change less than ±3 °C after self-calibration execution
5. Calibration period: 1 year
6. Measurement speed: 1 PLC (Power Line Cycle)
Maximum voltage and current
Max voltage Max current
DC or pulsed
1
210 V 0.105 A
21 V 1.515 A
2
6 V 3.03 A
2
Pulsed only
1
200 V 1.515 A
6 V 10.5 A
1. See “Maximum pulse width and duty cycle” in Pulse Source Supplemental Characteristics for
applicable maximum voltage and current.
2. Max current limitation: For 21 V/1.515 A and 6 V/3.03 A ranges, total max current is limited by the
table below for using 2 channels. Max current is not limited for using 1 channel only.
Maximum current limitation
Ch1 voltage Ch2 voltage Max total current limitation of Ch1 and Ch2
-200
Voltage [V]
Pulsed only
1
-10.5
-6 6
200
1.515
10.5
-1.515
3.03
Voltage [V]
21
-21
-6
-0.1
6
-1.515
0.1
1.515
-3.03
210
DC or pulsed
1
04 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet