Module 5
Basic Measurement
5-46
z Use single triax for basic MOS FET measurements
z One triax on each Force line
z Leave the Sense lines open...no cable connected
z Use triax bulkhead feedthru adapters on the prober
z Use a probe holder with a triaxial connector
z Use a probe holder than accommodates replaceable
probe tips (inexpensive repair)
Simplifying the Prober Connection
The next section sheds light on the confusing topic of connection to the device under test. Without
proper considerations, noise, capacitance, sense resistance, or inductance will result in unacceptable
measurement error.