Chapter 6: Replacing Assemblies
To remove and replace the bottom sleeve
100
8
Remove the two T20 handle screws to remove the handle.
9 Gently lift the bottom sleeve up and out. Be careful to not catch the feet on the chassis.
Figure 6-3
Removing bottom sleeve
10 To replace the bottom sleeve, reverse the above procedure.
Be sure to keep ribbon cables out of the way when replacing the cover, particularly the flex cable
and connector for the AutoProbe assembly at the bottom front of the oscilloscope.