To save or recall traces
With the Measurement/Storage Module installed, the two volatile pixel
memories are replaced with four high-speed non-volatile memories.
In addition, 64 Kbytes of nonvolatile trace memory with data compression is
also provided. A data compression algorithm maximizes the number of traces
and front-panel setups that can be stored into this memory.
The total number of traces that can be saved depends on the complexity of
each trace. At least 4 highly complex traces, or up to 96 simple traces, can
be saved in this 64 Kbytes of memory. Storage time is less than 10 seconds.
1
Adjust Oscilloscope controls for desired stable display.
2 Press Trace .
A softkey menu with six choices appears. All of them are related to the trace
memories of the module.
Trace Selects the trace memory (ALL, 1 to 100). When ALL is selected,
four additional softkeys are displayed.
a All Off Used to turn all 100 trace memories to off.
b All On Used to turn all 100 trace memories to on.
c Clear All Used to clear the contents of all 100 trace memories at
one time.
d Review Traces Used to view the contents of all trace memories
(with stored data) one at a time. View time is approximately 3
seconds. When selected, two additional softkeys are displayed.
Pause Review pauses review cycle on current trace until key is
pressed again, and Cancel Review cancels the review cycle.
Trace X Off On Turns the selected memory number to on or off. When on
is selected, trace status is temporarily displayed on the screen.
Save to Trace X Saves the currently displayed waveform and front-panel
setup to the selected trace memory location.
Clear Trace X Erases a previously stored waveform and front-panel setup
from the selected trace memory location.