Agilent Technologies 4000 X Instrukcja Obsługi Strona 83

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 168
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 82
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs 4
4000 X-Series Oscilloscopes Advanced Training Guide 83
Note that your scope should still be triggering on random edge crossings of
channel- 1. Let’s now set up the scope to trigger when it detects a Read
operation from address = 29
HEX
, following by an “acknowledge”, followed
by any data content (don’t care).
18 Press the [Trigger] front panel key; then select Serial 1 (I
2
C) using the
Entry knob.
19 Tap the Trigger: Start softkey; then select Frame(Start:Addr 7:Read:Ack:Data).
20 Tap the Address softkey; then set to 0x29 using the Entry knob.
21 Set the scope’s timebase to 200.0 µs/div.
You should now see stable waveform traces on your scope’s display similar
to Figure 52. The I
2
C frame that is near the center of the scope’s display,
which is the frame that the scope is triggering on, should be decoded as:
< 29RA FFA 80~A > or < 29RA FFA C0~A >
Figure 52 Triggering on I
2
C traffic.