Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics
Trigger system
Trigger sources Analog channel (1~4), digital channel (D0~D15), line, external, WaveGen (1 or mod) (FM/FSK)
Trigger modes Normal (triggered): requires trigger event for scope to trigger
Auto: triggers automatically in absence of trigger event
Single: triggers only once on a trigger event, press [Single] again for scope to find another
trigger event, or press [Run] to trigger continuously in either Auto or Normal mode
Force: front panel button that forces a trigger
Trigger coupling DC: DC coupled trigger
AC: AC coupled trigger, cutoff frequency: < 10 Hz (internal); <50 Hz (external)
HF Reject: High frequency reject, cutoff frequency ~ 50 kHz
LF Reject: Low frequency reject, cutoff frequency ~ 50 kHz
Noise Reject: Selectable OFF or ON, decreases sensitivity 2x
Trigger holdoff range 40 ns to 10.00 s
Trigger sensitivity
Internal* < 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
External* 200 mVpp from DC to 100 MHz
350 mVpp 100 MHz to 200 MHz
Trigger level range
Any channel ± 6 div from center screen
External ± 8 V
* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.