M3000A/M3046A/M3015A/M3016A Service Guide
6-8 Repairing the Instrument
6WHS Push the complete assembly (including the rubber cushion, the LCD display, and
the cables) out from the rear of the chassis. Start pushing from the top being careful
not to damage or bend the 6 clips around the cushion:
&$87,21
Never touch the front of the display assembly (the screen), or the backlight tubes, with your
fingers. Doing so can reduce the lifetime of the parts.
6WHS Then push from the sides:
6WHS Lay the LCD display on a flat, scratch-free surface.
6WHS Unplug the backlight tube cables.