Agilent Technologies E6640A EXM Instrukcja Obsługi Strona 27

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 26
Quick-Start 1
Getting Started Guide 27
To define RF4 I|O as an input and RF3 I|O as an output, use the following steps.
Define RF4 I|O as an input, using this menu setting or its equivalent command:
Input/Output > More > RF Output & Test Set Config > Half Duplex Config >
RF Input Port > RF4 I|O
Command:
:HDUPlex:PORT:INPut RFIO4
Define RF3 I|O as an output, using this menu setting or its equivalent command:
Input/Output > More > RF Output & Test Set Config > Half Duplex Config > RF
Output Port > RF3 I|O
Command:
[:SENSe]:HDUPlex:PORT:OUTPut RFIO3
NOTE This does not complete the port setup; two more settings are required!
Set up the RF Input using this menu setting or its equivalent command:
Input/Output > RF Input > RF Input Port > RF Input
Command:
:FEED:RF:PORT RFIN
Set up the RF Output using this menu setting or its equivalent command:
Input/Output > More > RF Output & Test Set Config > RF Output Port > RF Output
Command:
:FEED:RF:PORT:OUTP RFO
Przeglądanie stron 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 117 118

Komentarze do niniejszej Instrukcji

Brak uwag