Agilent Technologies DP235 Podręcznik Użytkownika Strona 312

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 602
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 311
Contents 312 Index
Performing Qualifications
To ensure a validated Agilent 2100 bioanalyzer system, qualification steps have to be
performed at installation and operation level.
2100 expert allows for detailed installation qualification (IQ) and operational qualification
(OQ) on both the bioanalyzer hardware and software. Each qualification comprises a
series of tests and measurements that you can run and document in the Validation
context of the 2100 expert software.
Installation Qualification
Installation qualification includes tests to verify that the bioanalyzer software and
hardware are installed properly and that all electrical and pressure connections are
correct.
Installation qualification must be performed once after installation.
Przeglądanie stron 311
1 2 ... 307 308 309 310 311 312 313 314 315 316 317 ... 601 602

Komentarze do niniejszej Instrukcji

Brak uwag