Source Rise And Fall Time Mismatch Tests 11
U7232A DisplayPort Compliance Testing Notes 109
PASS Condition
Falling Mismatch ≤ 15% of the single- ended rise time
Rising Mismatch ≥ 15% of the single- ended fall time
Test R ef er en ce s
See section 3.7, in the DisplayPort- Compliance Test Specification Version
1.
Rise-Fall Mismatch Edge Sets the number of edges measured for the rise-fall mismatch test.
Upper Threshold (Debug
only)
Sets the upper threshold used to make a rise time or fall time
measurement.
Lower Threshold (Debug
only)
Sets the lower threshold used to make a rise time or fall time
measurement.
Intra-pair Skew
Intra-pair Edge Sets the number of edges measured for the intra-pair skew test.
AC Common Mode
AC Common Edge Sets the number of edges measured for the AC common mode test.
Table 12 Test Configuration Options