Using Microsoft Windows XP
Embedded Web Server: Using the X-Series Instruments Remotely
Selecting the SCPI Telnet tab
Selecting the SCPI Telnet tab opens a Telnet session between the instrument and the
client computer. The instrument TCP/IP port used for SCPI access is 5023.
NOTE To run a SCPI Telnet session, the instrument application must be running.
Step Notes
1. Click SCPI Telnet A password entry dialog appears.
By default, this password is set at the factory
as “agilent”. However, you may subsequently
change the password. (Press System, I/O
Config, Reset Web Password on the
instrument front panel to change the
password.)
When the correct password is entered, the
connection window appears.
2. Click SCPI Telnet Session The telnet command line interface appears.