Agilent Technologies E8267D PSG Instrukcja Użytkownika Strona 14

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 24
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 13
14
External I/Q inputs
Input impedance switched 50 or 600 (nominal)
Input range
1
Minimum 0.1 V
rms
, maximum 1V
peak
Flatness ± 1 dB within ± 40 MHz of carrier (with ALC off) (typical)
Measured I/Q frequency response
2
I/Q adjustments
I & Q offsets External inputs (600 ) ± 5 Volts
External inputs (50 ) ± 50 %
Internal baseband generator ± 50 %
I/Q attenuation 0 to 40 dB
I/Q gain balance ± 4 dB
I/Q quadrature skew ± 10° range (typical)
Low pass filter Selectable 40 MHz or through
I/Q baseband outputs
Differential I, I bar, Q, Q bar
Single ended I, Q
Frequency range DC to 40 MHz
Output voltage into 50 W 1.5 V
p-p
(typical)
DC offset adjustments ± 3 V
DC offset resolution 1 mV
Low pass filter Selectable 40 MHz or though
Vector modulation
Offset from carrier (GHz)
Conversion loss (dB)
5
0
-5
-10
-15
-0.15 -0.10 -0.05 0.00 0.05 0.10 0.15
Offset from carrier (GHz)
Conversion loss (dB)
5
0
-5
-10
-15
-0.15 -0.10 -0.05 0.00 0.05 0.10 0.15
Offset from carrier (GHz)
Conversion loss (dB)
5
0
-5
-10
-15
-0.15 -0.10 -0.05 0.00 0.05 0.10 0.15
Offset from carrier (GHz)
Conversion loss (dB)
5
0
-5
-10
-15
-0.15 -0.10 -0.05 0.00 0.05 0.10 0.15
1.5 GHz 3 GHz
12 GHz 20 GHz
1. For optimum signal quality, the I and Q inputs should be 0.7 V
peak
, with (I
2
+ Q
2
) + 150 mV
rms
. Different
RMS levels are accommodated by adjusting the internal I/Q modulator attenuator, which may be either
manually or automatically set. The minimum input level required to maintain RF level accuracy is
(I
2
+ Q
2
) = 0.1 V
rms
.
2. Sine wave response, measured with input level = 100 mV
rms
on one channel, and ALC off.