Error Messages
Introduction
Introduction
If an error condition occurs in the signal generator, it is reported to both the front panel display error queue
and the SCPI (remote interface) error queue. These two queues are viewed and managed separately.
NOTE When there is an unviewed message in the front panel error queue, the ERR annunciator
appears on the signal generator’s display.
Characteristic Front Panel Display
Error Queue
SCPI Remote Interface
Error Queue
Capacity (#errors) 30 30
Overflow Handling Circular (rotating).
Drops oldest error as new error comes in.
Linear, first-in/first-out.
Replaces newest error with:
-350, Queue overflow
Viewing Entries Press:
Utility > Error Info > View Next
(or
Previous) Error Message
Use SCPI query
SYSTem:ERRor[:NEXT]?
Clearing the Queue Press:
Utility > Error Info >
ClearErrorQueue(s)
Power up
Send a *CLS command
Read last item in the queue
Unresolved Errors
(errors that must be
resolved.
For example: unlock,
avalanched, and
hi/lo)
Re-reported after queue is cleared. Re-reported after queue is
cleared.