Agilent Technologies E4438C Podręcznik Użytkownika Strona 24

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 34
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 23
16 User’s and Service Guide Supplement
E4438C Option HEC
Repeatable Triggering for Multiple Baseband Sources
3. Select ARB Setup > More (1 of 3) > More (2 of 3) > VCO CLOCK Ext > [Return] ARB
“On. Refer to Figure 11.
4. Adjust the trigger level control on the oscilloscope for a stable display.
5. Align the DACs:
[Mode] Dual ARB > ARB Setup > More (1 of 3) > More (2 of 3) > Align DAC
NOTE
Changing or disrupting the external clock frequency can cause the DACs to
become misaligned. Press the align DACs softkey after changing the clock
frequency.
Figure 11 I & Q Output for the “Master” Instrument
Przeglądanie stron 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 33 34

Komentarze do niniejszej Instrukcji

Brak uwag