Agilent Technologies E4438C Przewodnik Instalacji Strona 109

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 140
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 108
Adjustments
FM In-Band DAC Offset Calibration
4-17
FM In-Band DAC Offset Calibration
Description
This calibration is used to remove the offset associated with the FM IN-BAND DAC located on the A17
Synthesizer (Option 501, 502, 503, 504). The calibration determines the DAC value for the FM IN-BAND
OFFSET DAC on the A17 Synthesizer (Option 501, 502, 503, 504) which will remove the offset.
Required Test Equipment
Agilent 53132A Option 050 frequency counter
Procedure
Figure 4-11 FM In-Band DAC Offset Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controllers display.
Przeglądanie stron 108
1 2 ... 104 105 106 107 108 109 110 111 112 113 114 ... 139 140

Komentarze do niniejszej Instrukcji

Brak uwag