Troubleshooting
Troubleshooting Adjustment Problems
1-87
Power Level Accuracy, Low Power
Calibration
AT1 High-Power Mechanical Attenuator (Option UNB)
A28 Reverse Power Protection (Option 501, 502, 503, 504
or Option UNJ)
Loose or leaky RF cables.
Power Search Calibration A13 Output
A14 Extended Frequency Output (Option 506)
AM Gain Calibration A13 Output
A14 Extended Frequency Output (Option 506)
I/Q Gain/Offset/Quadrature Calibration A10 I/Q Multiplexer
A13 Output (Option 501, 502, 503, 504 or Option UNJ)
A14 Extended Frequency Output (Option 506)
A7 Baseband Generator (Option 001 or Option 002)
I/Q Impairment Calibration A10 I/Q Multiplexer
A13 Output
A14 Extended Frequency Output (Option 506)
Table 1-39 Troubleshooting Adjustment Problems
Adjustment that is failing...
The following adjustments are listed in
the order that they should be performed
for proper calibration.
Assemblies most likely causing failure...