Subsystem Syntax STATus
:OPC
:INITiate <
state>
:OPERation
:CONDition? [query only]
:ENABle <
unmask>
[:EVENt]? [query only]
:NTRansition <
unmask>
:PTRansition <
unmask>
:PRESet [no query]
:QUEStionable
:CONDition? [query only]
:ENABle <
unmask>
[:EVENt]? [query only]
:NTRansition <
unmask>
:PTRansition <
unmask>
:OPC:INITiate
STATus:OPC:INITiate <state> controls whether the *OPC, *OPC?, and *WAI
commands will complete immediately or whether they will wait for waveform
generation to complete. With state
OFF set, these commands will complete
immediately. With state
ON set, they will wait for the Pending Operation Flag set
true by
INITiate:IMMediate to return false, indicating that the trigger system is in the
idle state and that waveform generation has completed or been aborted by the
ABORt or *RST commands.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
state>
boolean OFF | 0 | ON | 1 none
Comments Executable when Initiated: Yes
Coupling Group: None
Related Commands: *OPC, *OPC?, *RST, *WAI, ABORt, INITiate:IMMediate,
STATus:PRESet
*RST Condition: Unaffected
Power-on Condition: STATus:OPC:INITiate ON
Example Setting Immediate Completion Mode
STAT:OPC:INIT OFF Completes immediately for *OPC, etc.