Specifications (continued)
Zero and measurement noise
1. Condition: (i) 0 °C to 55 °C and (ii) 95% relative humidity at 40 °C non-condensing.
2. Within one hour after zero set, at a constant temperature, after a 24-hour warm-up of the power sensor.
3. The number of averages at 16 for Normal mode, measured over a one-minute interval and two standard deviations.
4. Noise per sample specifications are not applicable to U2004A. They are only applicable for gated power working range
as stated in the ‘Gate’ table on page 15.
Power range
1
U2000/1/2/4A sensors
–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
U2000/1/2H sensors
–50 dBm to –25 dBm
–28 dBm to –5 dBm
–10 dBm to 1 dBm
–1 dBm to 5 dBm
3 dBm to 25 dBm
20 dBm to 30 dBm
U2000/1B sensors
–30 dBm to –5 dBm
–8 dBm to 15 dBm
10 dBm to 21 dBm
19 dBm to 25 dBm
23 dBm to 44 dBm
No. of averages
Noise multiplier
Normal mode
x2 mode
1
2.0
2.7
2
1.8
2.4
4
1.7
2.0
8
1.5
1.6
16
1.0
1.0
32
0.95
0.91
64
0.74
0.78
128
0.55
0.53
256
0.39
0.34
512
0.29
0.29
1024
0.21
0.20
Zero set (internal)
±1.5 nW (±2.8 nW for U2004A)
±2 nW (±3 nW for U2004A)
±12 nW
±2 μW
±4 μW
±6 μW
±15 nW
±20 nW
±120 nW
±20 μW
±40 μW
±60 μW
±1.8 μW
±2 μW
±12 μW
±2 mW
±4 mW
Zero set (external)
±600 pW
±1.5 nW
±10 nW
±500 nW
±1 μW
±5 μW
±8 nW
±20 nW
±100 nW
±20 μW
±30 μW
±60 μW
±800 nW
±2 μW
±10 μW
±1 mW
±2 mW
Zero drift
2
200 pW
400 pW
1.5 nW
50 nW
500 nW
2 μW
2 nW
4 nW
15 nW
500 nW
5 μW
20 μW
200 nW
400 nW
1.5 μW
50 nW
500 μW
Measurement noise
3
1 nW
1.5 nW
15 nW
650 nW
1 μW
10 μW
10 nW
15 nW
150 nW
6.5 μW
10 μW
100 μW
1 μW
1.5 μW
15 μW
650 μW
1 mW
Noise Per Sample
4
N/A
90 nW
1 μW
55 μW
85 μW
550 μW
N/A
900 nW
10 μW
550 μW
850 μW
5.5 mW
N/A
90 μW
1 mW
55 mW
85 mW
Effects of averaging on noise: Averaging over
1 to 1024 readings is available for reducing
noise. The table below provides the
measurement noise for a particular sensor with
the number of averages set at 16 (for normal
mode) and 32 (for x2 mode). Use the noise
multiplier, for the appropriate of averages, to
determine the total measurement noise value.
Example:
U2000A power sensor, –60 dBm to –35 dBm,
normal mode, number of averages = 4
Measurement noise calculation:
1 nW x 1.7 = 1.7 nW
12