8753ET Specifications and Characteristics
Instrument Specifications
Table 2-15 Test Port Input
8753ET R-Channel Input
Dynamic Accuracy (Typical)
Accuracy of the R-channel power reading relative to the R-channel reference power level.
This chart applies when the calibrated power level and the measurement power level are not the
same.
Inputs: R-channel
300 KHz to 3 GHz 3 to 6 GHz
Magnitude Dynamic Accuracy
for Test Port Powers
> 50 dBm and < 0 dBm .02 dB + .001 dB/dB
from the reference power
.02 dB + .001 dB/dB
from the reference power
> 0 dBm and < 10 dBm .02 dB + .02 dB/dB
above 0 dBm
.02 dB + .03 dB/dB
above 0 dBm
Phase Dynamic Accuracy for
Test Port Powers
> 50 dBm and < 0 dBm .132 deg + .0066 deg/dB
from the reference power
.132 deg + .0066 deg/dB
from the reference power
> 0 dBm and < 10 dBm .132 deg + .132 deg/dB
above 0 dBm
.132 deg + .198 deg/dB
above 0 dBm