16
□ High-power test set (Option H85)
This configuration combines options that are often necessary
for high power measurements (UNL*
1
, 014, 016, 080, 081). The
only difference between ordering Option H85 versus a
combination of the options listed above is the source
attenuator option UNL. Standard UNL includes two source
attenuators and two bias-tees. Option H85 includes the two
source attenuators, but not the bias-tees, as the bias-tees are
the power-limiting factor in the network analyzer test set.
The maximum power at the test port is +43 dBm (<20 GHz), and
+40 dBm (>20 GHz).
Option 080, frequency-offset mode, is included in Option H85
because it manages the phase-locking internally (instead of
depending on the R1 receiver). So if you need to use external
components in the path of the R1 receiver, it makes the
measurements simpler and more robust.
□ Rack mount kit without handles (Option 1CM)
Adds a rack mount (5063-9217) and rail kit (E3663AC) for use
without handles.
□ Rack mount kit with handles (Option 1CP)
Adds a rack mount (5063-9237)
3
and rail kit (E3663AC) for use
with standard supplied handles.
Documentation
PNA Online Help system is available within PNA instruments in
the following languages: English, German, Spanish, French,
Japanese, and Chinese.
The PNA Service Guide and Online Help are available on the
Web: http://na.tm.agilent.com/pna
□ Add receiver attenuators (Option 016)
An attenuator is added between each test port and its
corresponding receiver. Two 35 dB step attenuators are added
to the E8362/3/4B. Two 50 dB step attenuators are added to
the E8361A (see PNA Series Microwave Data Sheet literature
number 5988-7988EN for a basic block diagram).
□ Pulsed-RF measurement capability (Option H08)
2
Provides software to set up and control pulsed-RF measurements
with point-in-pulse capability. The software sets the coefficient
of the PNA’s digital-IF filter to null out unwanted spectral
components, enables the IF gates provided with IF access
(Option H11), and controls selected Agilent pulse generators.
It can be run on the PNA or an external computer. A “.dll” file
containing the IF-filter algorithms is included for automated
pulsed-RF testing. The pulsed application is configured to work
with the Agilent 81110A series pulse generator.
For more detailed information regarding pulsed measurement
capabilities with the microwave PNA refer to the Agilent Web
site www.agilent.com/find/pna and download the PNA Series
MW Network Analyzers Configuration Guide for Pulsed
Measurements, literature number 5988-9833EN.
□ IF access (Option H11)
Provides hardware to enable antenna, point-in pulse, and
broadband millimeter-wave measurements to 110 GHz. For
each of the MW PNA’s measurement receivers, IF gates
(enabled with pulsed measurement capability, Option H08) and
external IF inputs are added. In addition, access to the PNA’s
internal RF and LO source is provided for remote mixing
applications. For basic antenna measurements, only Option H11
is necessary. Pulsed antenna applications also require the
pulsed measurement capability (Option H08). Broadband
measurements to 110 GHz, also requires an N5260A
millimeter-wave test set controller.
Note: Use external IF access for up to 20 dB more sensitivity
when making antenna measurements with a remote mixing
configuration. Add Option H08 (Pulsed-RF Measurement
Capability) to enable advanced pulsed measurements. Or
upgrade to a broadband (10 MHz to 110 GHz) VNA system
simply by purchasing an N5260A controller test set with test
heads (Option 110, 120, or 130).
1. UNL* does not include bias-tees. Only includes source attenuators.
2. Up to 67 GHz.
3. The 5063-9237 kit assumes you have the standard handles shipped
with the instrument. If you do not have handles, order a 5063-9224 kit.