Agilent Technologies U1270 Series Instrukcja Użytkownika Strona 3

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 21
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 2
3
Figure2.U1272Ahelpsyouidentifythepresenceofstrayvoltageonadisconnectedwirerunningparallelwiththewire
poweringuptheVFDtoanindustrialmotor.TheimageontherightshowstheU1272Ainlowimpedancemode.
Low impedance mode
Stray voltages are usually found
in non-energized electrical wiring
adjacent to powered wires due to
capacitive or inductive coupling
between these wires. The low
impedance mode serves to eliminate
false readings by dissipating these
stray voltages thus improves safety
and measurement efficiency during
voltage measurement.
Peak detect at 250 µs
The peak detect function allows
you to capture the engine or motor
startup transient as fast as 250 µs.
Key Functions