Safety
Design
The system is designed to prevent
electrical, mechanical, or radiation
hazards to the operator. The
system is fully certifi ed to meet
all U.S. Federal specifi cations
regarding the use of cabinet X-ray
systems in an industrial environ-
ment. All design requirements
meet or exceed 21 CFR 1020.40,
subchapter J. The system meets
the applicable international
government radiation emission
standards (including German
and Japanese standards).
Safety interlocks
Redundant safety interlocks are
used to eliminate the possibility
of accidental X-ray exposure or
physical harm.
Allowable panel characteristics
Panel size
Panel size
Maximum
Maximum
2
2
457 mm x 609 mm (18.0 in x 24.0 in)
457 mm x 609 mm (18.0 in x 24.0 in)
(width x length)
(width x length)
1
1
Minimum
Minimum
3
3
102 mm x 127 mm (4.0 in x 5.0 in)
102 mm x 127 mm (4.0 in x 5.0 in)
Aspect ratio Length ≥ 0.5 x panel width
Aspect ratio Length ≥ 0.5 x panel width
Maximum panel 445 mm x 609 mm (17.5 in x 24 in)
Maximum panel 445 mm x 609 mm (17.5 in x 24 in)
inspectable area
inspectable area
Panel thickness
Panel thickness
Maximum 3.2 mm (0.125 in)
Maximum 3.2 mm (0.125 in)
Minimum
Minimum
4
4
0.5 mm (0.020 in)
0.5 mm (0.020 in)
Panel clearance
Panel clearance
T
T
op clearance
op clearance
5
5
25.0 mm (1.0 in)
25.0 mm (1.0 in)
Bottom clearance
Bottom clearance
5
5
50.0 mm (2.0 in)
50.0 mm (2.0 in)
Panel edge clearance 3.0 mm (0.118 in) on parallel edges of the panel
Panel edge clearance 3.0 mm (0.118 in) on parallel edges of the panel
Panel width tolerance ±0.7 mm (0.0275 in)
Panel width tolerance ±0.7 mm (0.0275 in)
Panel weight
Panel weight
2
2
Maximum 4.5 kg (10.0 lb)
Maximum 4.5 kg (10.0 lb)
Minimum 0.03 kg (0.066 lb)
Minimum 0.03 kg (0.066 lb)
Maximum acceptable
Maximum acceptable
40ºC (104ºF)
40ºC (104ºF)
panel temperatures
panel temperatures
(at time of load)
(at time of load)
Panel warp
Panel warp
Warp of less than 1% per unit of linear measurement.
Warp of less than 1% per unit of linear measurement.
(after refl ow
(after refl ow
and
and
Maximum downside panel warp cannot exceed 2.0 mm (0.08 in).
Maximum downside panel warp cannot exceed 2.0 mm (0.08 in).
wave soldering)
wave soldering)
Maximum upside panel warp cannot exceed 1.0 mm (0.04 in).
Maximum upside panel warp cannot exceed 1.0 mm (0.04 in).
Maximum number
Maximum number
No restriction
No restriction
of
of
PCBs per panel
PCBs per panel
1
1
Panels are handled on width edges. Panels with edge cut outs may require the use of a carrier.
Panels are handled on width edges. Panels with edge cut outs may require the use of a carrier.
2
2
Maximum panel size dimensions and weight must include carrier if applicable.
Maximum panel size dimensions and weight must include carrier if applicable.
3
3
Smaller panels are possible with the use of panel carriers. A carrier can only have one panel.
Smaller panels are possible with the use of panel carriers. A carrier can only have one panel.
4
4
With panels of this thickness, imaging results can be affected by PCBA layout.
With panels of this thickness, imaging results can be affected by PCBA layout.
5
5
Measured from the bottom of the panel including a maximum warp.
Measured from the bottom of the panel including a maximum warp.
Power and environmental
Voltage requirements
Voltage requirements
200-240 VAC three phase; 380-415 VAC three phase wye (±5%)
200-240 VAC three phase; 380-415 VAC three phase wye (±5%)
(50 Hz or 60 Hz)
(50 Hz or 60 Hz)
Total harmonic
Total harmonic
THD <5%, single harmonic <3%
THD <5%, single harmonic <3%
distortion (THD)
distortion (THD)
Ground to
Ground to
Voltage <1 VRMS
Voltage <1 VRMS
neutral voltage
neutral voltage
Momentary voltage
Momentary voltage
Interruptions, defi ned as: voltage drop on one or more phase
Interruptions, defi ned as: voltage drop on one or more phase
conductors below 180 VAC for greater than 10 ms
conductors below 180 VAC for greater than 10 ms
Primary power rating
Primary power rating
2.8 kVA
2.8 kVA
Maximum inrush
Maximum inrush
58 amps RMS
58 amps RMS
current
current
Circuit breaker rating
Circuit breaker rating
10,000 amps interrupting capacity (AIC) for voltages up to 240 VAC,
10,000 amps interrupting capacity (AIC) for voltages up to 240 VAC,
14,000 AIC for voltages greater than 240 VAC
14,000 AIC for voltages greater than 240 VAC
Wiring requirements
Wiring requirements
Reference the Agilent
Reference the Agilent
Medalist
Medalist
x6000 Installation Guide for hookup
x6000 Installation Guide for hookup
wiring confi gurations, full load amperes, circuit breaker size/#
wiring confi gurations, full load amperes, circuit breaker size/#
poles and conductor size
poles and conductor size
Note:
Note:
local electrical wiring codes take precedence
local electrical wiring codes take precedence
Air requirements
Air requirements
552 kPa (80 psi) compressed air
552 kPa (80 psi) compressed air
Temperature
Temperature
16-30ºC, 60-86ºF
16-30ºC, 60-86ºF
requirements
requirements
Humidity requirements
Humidity requirements
Relative humidity 20-60%
Relative humidity 20-60%