Agilent Technologies ENA Series Instrukcja Obsługi Strona 38

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 37
38 Chapter 2
Performance Tests
E5061A Performance Test Record (for Option 150/250)
Level Accuracy Test (@ Port 2) (Option 250 only)
Level Flatness Test (@ Port 2) (at 0dBm, relative to 50 MHz reference) (Option 250
only)
RF Output Level Linearity Test (@ Port1)
CW Frequency: 300 kHz (relative to 0 dBm reference)
1.5 G 1.0 0.08
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level
[dBm]
Frequency
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
0 50 M 0.8 0.07
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
300 k 1.0 0.27
1.5 M 1.0 0.07
3.00 M 1.0 0.07
7.00 M 1.0 0.07
15.0 M 1.0 0.07
275 M 1.0 0.07
575 M 1.0 0.07
875 M 1.0 0.07
1.175 G 1.0 0.07
1.5 G 1.0 0.08
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 0.75 0.03
-2.5 0.75 0.03
2.5 0.75 0.03
5 0.75 0.03
7.5 0.75 0.03
Przeglądanie stron 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 250 251