280 Chapter 9
Performance Test Records
CDMA Analyzer Channel Power Level Accuracy Performance Test 28 Record
CDMA Analyzer Channel Power Level Accuracy
Performance Test 28 Record
For test procedure, see “CDMA Analyzer Channel Power Level
Accuracy Performance Test 28” on page 227.
Table 9-29 CDMA Analyzer Channel Power Level Accuracy
Test 28 Record
RF
(MHz)
Level
(dBm)
Measured Level Limits (dBm)
Lower Upper Actual
881.52 11 10.25 11.75
881.52 6 5.25 6.75
881.52 1 0.25 1.75
881.52
4 4.75 3.25
881.52 9 9.75 8.25
881.52
14 14.75 13.25
881.52
19 19.75 18.25
1931.25 11 10.25 11.75
1931.25 6 5.25 6.75
1931.25 1 0.25 1.75
1931.25
4 4.75 3.25
1931.25
9 9.75 8.25
1931.25
14 14.75 13.25
1931.25
19 19.75 18.25