Agilent Technologies 83630A Dokumentacja Strona 120

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 119
Second Harmonic Distortion Troubleshooting
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Second Harmonic Distortion Troubleshooting
Related Adjustments
Lowband Mixer Bias Adjustment
Switched LO Distribution Amplifier Adjustment
Failures when the source frequency is less than or equal to 1.5 GHz are usually a low
band path problem. Suspect the A20 Lowband Assembly or something between the input
connector and the Lowband assembly.
Failures when the source frequency is greater than 1.5 GHz are usually a high band
problem. Suspect the A19 RYTHM or something between the input connector and the
RHYTHM assembly. See PSA Service Guide for more troubleshooting information.
http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Troubleshooting/troubleshd.htm [5/29/2003 8:47:44 AM]
Przeglądanie stron 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 132 133

Komentarze do niniejszej Instrukcji

Brak uwag