53220A/53230A User’s Guide 235
53220A/53230A Math, Graphing, and Data Logging 6
Pressing View Results after logging Time Stamp measurements will display
“Graphics not supported for this function”.
Figure 6-12 is an example of a trend chart as readings are logged. (Figures 6-8
and 6-10 contain more information on the trend chart window.)
Figure 6-12. Trend Chart Display During Data Logging.
When logging is complete, the Marker, Pan, and Zoom softkeys can be used
review the data. Note that pressing the Graph key followed by the Trend Chart
softkey does not retain the data in the display.
If greater than 100 readings are logged the readings are decimated - grouped
together and represented by decimation points (see “Trend Charts” and
“Reading Decimation”).
Stopping the Application
Pressing Stop Save or Stop No Save stops data logging before the duration (time
or number of readings) is reached. All readings logged to that point are saved
or discarded depending on the key pressed.
Function
Channel
Current Reading Count
(memory)
time or number of reading
s
remaining
reading indices
T
rend Window
(100 points)