Agilent Technologies 437B Podręcznik Użytkownika Strona 105

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 261
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 104
General Power Meter Functions 2
N1913/1914A EPM Series Power Meters User’s Guide 83
Agilent 8485D Prior to the power meter being zeroed and calibrated, an Agilent
11708A 30 dB reference attenuator and an APC 3.5 (f) to 50 Ω (m)
N-Type adapter (08485-60005) should be connected between the
power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
Agilent 8487A This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) to connect to the power meter. Remove this attenuator
before making measurements.
Agilent 8487D Prior to the power meter being zeroed and calibrated, an Agilent
11708A 30 dB reference attenuator and an APC 2.4 (f) to 50 Ω (m)
N-Type adapter (08487-60001) should be connected between the
power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
Sensor Model Connection Requirements
Przeglądanie stron 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 260 261

Komentarze do niniejszej Instrukcji

Brak uwag