Strona 1 - Multimeter

Keysight 3458A MultimeterCalibration Manual

Strona 2

8 Chapter 1 3458A Calibration Introduction is shipped from the factory with its security code set to 3458. Specifying 0 for the new_code in the SECURE

Strona 3

98 Appendix A Specifications

Strona 4

Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3) 99Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3)A v

Strona 5 - Documentation History

100 Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3) Saving Calibration Time and MoneyThe increasing accuracy required of today

Strona 6 - DECLARATION OF CONFORMITY

Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3) 101Measurements using a Josephson junction standard confirm linearity of the

Strona 7 - Contents

102 Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3) The user enters the exact value of the external 10 V DC voltage standard (

Strona 8 - 6 Contents

Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3) 103Resistance and DC Current AdjustmentsCalibration of all resistance (nine ra

Strona 9 - Calibration Security

104 Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3) The 3458A Multimeter produces the required step input voltage. Then, its a

Strona 10 - Security Code

Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3) 105SummaryElectronic internal calibration of the 3458A Multimeter simplifies

Strona 11 - Calibrations

106 Appendix B Electronic Calibration of the 3458A (Product Note 3458A-3)

Strona 12 - Violations

This information is subject to change without notice. © Keysight Technologies 1988 - 2014 Edition 7, August 2014*03458-90017*03458-90017www.keysight.c

Strona 13 - Non-Destructive Input

Chapter 1 3458A Calibration Introduction 9display the SECURE command).f. Enter the number 0 followed by the delimiter (,) and the security code you

Strona 14

10 Chapter 1 3458A Calibration Introduction CAL. (or ACAL if secured) is executed. If autocal is secured, the calibration number is also incremented b

Strona 15 - Operational Tests

Chapter 1 3458A Calibration Introduction 11The remaining constants (such as internal reference and offset constants) are externally derived and not

Strona 16 - Function Offset Test

12 Chapter 1 3458A Calibration Introduction

Strona 17 - Gain Test

Chapter 2 Operational Verification Tests 13Chapter 2 Operational Verification TestsIntroductionThis section contains Operational Verification Tes

Strona 18

14 Chapter 2 Operational Verification Tests → scroll keys to view entire CAL? message. Record the temperatures on the Test Card.7. If the instrument

Strona 19 - Offset Test

Chapter 2 Operational Verification Tests 159. Remove the short from the rear input terminals.4-Wire OhmsFunction Gain TestThe following procedure ve

Strona 20

16 Chapter 2 Operational Verification Tests chapter 4, Performance Verification Tests.1. Execute the ACAL DCV command using the front panel "Aut

Strona 21

Chapter 2 Operational Verification Tests 17Chapter 3 "Adjustment Procedures" to make adjustments.DC Voltage FunctionOffset TestThis proced

Strona 23 - Required Equipment

18 Chapter 2 Operational Verification Tests

Strona 24

19OPERATIONAL TEST CARD - 1 YEAR LIMITSKeysight Model 3458A Digital Multimeter Test Performed BySerial Number DateCAL? 60TEMP?Difference - (must be l

Strona 26 - DC Gain Adjustment

Chapter 3 Adjustment Procedures 21Chapter 3 Adjustment ProceduresIntroductionThis section contains procedures for adjusting the 3458A Multimeter. T

Strona 27

22 Chapter 3 Adjustment Procedures The resultant accuracy of the multimeter depends on the accuracy of the equipment used, the thermal characteristic

Strona 28 - Current Adjustment

Chapter 3 Adjustment Procedures 23Figure 1. 4-Terminal ShortFront Terminal Offset AdjustmentThis adjustment uses an external 4-terminal short. The

Strona 29 - AC Adjustment

24 Chapter 3 Adjustment Procedures Note Take precautions to prevent thermal changes near the 4-wire short. You should not touch the short after it i

Strona 30

Chapter 3 Adjustment Procedures 25standard).Note Voltage standards from 1V DC to 12V DC can be used for this procedure. However, using a voltage st

Strona 31

26 Chapter 3 Adjustment Procedures Resistance and DCCurrent AdjustmentThis adjustment calculates gain corrections for the resistance and DC current r

Strona 32

Chapter 3 Adjustment Procedures 273. Set the front panel Terminals switch to Front.4. Connect the resistance standard to the multimeter's front

Strona 34 - 3458A Adjustment Record

28 Chapter 3 Adjustment Procedures the frequency and period measurement functions: adjusts the attenuator and amplifier high frequency response; and a

Strona 35

Chapter 3 Adjustment Procedures 293. Set the synthesizer to deliver a 3V rms sinewave at a frequency of 100 kHz. Connect the synthesizer, the 3V th

Strona 36

30 Chapter 3 Adjustment Procedures 8. Set the synthesizer to deliver a 100mV rms sinewave at a frequency of 100 kHz. Replace the 1V thermal converter

Strona 37

Chapter 3 Adjustment Procedures 3114. Repeat step 13 for each synthesizer setting and SCAL command shown on Lines D through H on the Adjustment Reco

Strona 38 - Procedure

32 Chapter 3 Adjustment Procedures *Always perform the above adjustments numbered 1 through 4: adjustment number 5 is only required once every 2 year

Strona 39

Chapter 4 Performance Verification Tests 33Chapter 4 Performance Verification TestsIntroductionThis chapter contains performance tests designed t

Strona 40 - DC Voltage Performance Tests

34 Chapter 4 Performance Verification Tests Test Considerations This section discusses many of the major problems associated with low-level measureme

Strona 41

Chapter 4 Performance Verification Tests 35• Ground Loops: Ground loops arise when the multimeter and the circuit under test are grounded at physica

Strona 42

36 Chapter 4 Performance Verification Tests General TestProcedureThe following performance tests utilize a transfer standard DMM to precisely measure

Strona 43

Chapter 4 Performance Verification Tests 37Figure 6. General Test Procedure

Strona 44 - AC Voltage Test

2KEYSIGHT TECHNOLOGIES WARRANTY STATEMENTKEYSIGHT PRODUCT: 3458A Multimeter DURATION OF WARRANTY: 1 year1. Keysight Technologies warrants Keysight ha

Strona 45

38 Chapter 4 Performance Verification Tests DC Voltage Performance TestsRequired Equipment The following equipment or its equivalent is required for t

Strona 46 - DC Current Performance Tests

Chapter 4 Performance Verification Tests 39DC Voltage FunctionOffset TestThe following procedure tests the offset voltage specification with the inp

Strona 47 - DC Current Function

40 Chapter 4 Performance Verification Tests of the DMM.4. Set the range of the DMM to 100 mV. Wait five minutes for the DMM to thermally stabilize.5.

Strona 48 - Ohms Performance Tests

Chapter 4 Performance Verification Tests 4116. After all DC gain tests have been performed, calculate and record in the column provided. The differe

Strona 49

42 Chapter 4 Performance Verification Tests Analog AC Voltage Performance TestsRequired Equipment The following list of equipment is required to test

Strona 50 - (Rear Terminals)

Chapter 4 Performance Verification Tests 43connection can be made using shielded test leads terminated with dual banana plugs. Refer to the general

Strona 51

44 Chapter 4 Performance Verification Tests points.11. If any of the differences calculated are greater than the specifiedlimits, refer to Chapter 3,

Strona 52 - Accuracy Test

Chapter 4 Performance Verification Tests 45DC Current FunctionOffset TestThe following procedure tests the DC current offset specifications with the

Strona 53 - DC VOLTAGE TESTS

46 Chapter 4 Performance Verification Tests current source.7. Connect the DC current source to the 3458A under test HI and LOinput terminals.8. Execut

Strona 54

Chapter 4 Performance Verification Tests 47done, turn the instrument ON and allow it to warm up before proceeding.3. The internal temperature of the

Strona 55 - AC VOLTAGE TESTS

3Safety SymbolsAlternating current (AC)Instruction manual symbol affixed to product. Indicates that the user must refer to the manual for specific WAR

Strona 56

48 Chapter 4 Performance Verification Tests 4-Wire OhmsFunction Offset Test(Rear Terminals)This procedure performance verifies the rear terminal ohms

Strona 57 - DC CURRENT TESTS

Chapter 4 Performance Verification Tests 49a. Note - At 10 MΩ, leakage current from the Sense leads introduce additionaluncertainty in the measureme

Strona 58

50 Chapter 4 Performance Verification Tests 4. Configure the DMM under test as follows:-- FREQ-- Trig SGL-- FSOURCE ACDCV-- LEVEL 0,DCFrequency Count

Strona 59 - OHMS TESTS

51PERFORMANCE TEST CARD - 1 YEAR LIMITSTest Performed ByKeysight Model 3458A Digital Multimeter Serial Number DateDC VOLTAGE TESTSCAL? 59TEMP?Differen

Strona 61 - ±00.00100 MHz ±00.00100 MHz

53PERFORMANCE TEST CARD - 1 YEAR LIMITSTest Performed ByKeysight Model 3458A Digital Multimeter Serial NumberDateAC VOLTAGE TESTSPerform an ACAL ACTes

Strona 63 - Chapter 5 Command Summary

55PERFORMANCE TEST CARD - 1 YEAR LIMITSTest Performed ByKeysight Model 3458A Digital Multimeter Serial NumberDateDC CURRENT TESTSCAL? 60TEMP?Differenc

Strona 65

57PERFORMANCE TEST CARD - 1 YEAR LIMITSTest Performed ByKeysight Model 3458A Digital Multimeter Serial NumberDateOHMS TESTSCAL? 60TEMP?Difference - (

Strona 66

4Declares, that the productConforms with the following European Directives:The product herewith complies with the requirements of the Low Voltage Dire

Strona 68

59PERFORMANCE TEST CARD - 1 YEAR LIMITSTest Performed ByKeysight Model 3458A Digital Multimeter Serial NumberDateFREQUENCY TESTSTest # 3458A Input3458

Strona 70

Chapter 5 Command Summary 61Chapter 5 Command SummaryThis section provides an alphabetical summary of commands that are used in calibrating the 3

Strona 71

62 Chapter 5 Command Summary ACALDescription Autocal. Instructs the multimeter to perform one or all of its automatic calibrations.Syntax ACAL [type]

Strona 72

Chapter 5 Command Summary 63Example OUTPUT 722;"ACAL ALL,3458" !RUNS ALL AUTOCALS, SECURITY CODE IS 3458! (FACTORY SECURITY CODE SETTING)

Strona 73

64 Chapter 5 Command Summary CALDescription Calibration. Calibrates the internal 7V reference to an external 10V standard (CAL10) and does the equiva

Strona 74

Chapter 5 Command Summary 65CAL?Description Calibration query. Returns a string containing one of four values for the calibration constant specifie

Strona 75 - Syntax CALNUM?

66 Chapter 5 Command Summary 22 ohm zero rear 1023 ohm zero rear 10024 ohm zero rear 1 K25 ohm zero rear 10 K26 ohm zero rear 100 K27 ohm zero rear 1

Strona 76 - Examples CALSTR

Chapter 5 Command Summary 6767 dci zero rear 10 mA68 dci zero rear 100 mA -ACAL OHMS69 dci zero rear 1A70 dcv gain 100 mV71 dcv gain 1V72 dcv gain 1

Strona 77 - Syntax REV?

Contents 5ContentsChapter 1 3458A Calibration IntroductionIntroduction ...7Calibration Se

Strona 78

68 Chapter 5 Command Summary 107 low aperture108 high aperture109 high aperture slope .01 PLC110 high aperture slope .1 PLC111 high aperture null .01

Strona 79 - Examples Changing the Code

Chapter 5 Command Summary 69151 overload ohm ocomp 100152 overload ohm ocomp 1 K153 overload ohm ocomp 10 K154 overload ohm ocomp 100 K155 overload

Strona 80 - Disabling Security

70 Chapter 5 Command Summary 196 flatness dac 100 mV197 flatness dac 1V198 flatness dac 10V199 flatness dac 100V200 flatness dac 1 KV201 level dac dc

Strona 81 - Syntax TEMP?

Chapter 5 Command Summary 71Remarks Related Commands: ACAL, CAL, SCALExample The following two program examples query the calibration constants. T

Strona 82 - Syntax TEST

72 Chapter 5 Command Summary 60 PRINT "CONST =", N70 OUTPUT 722;"QFORMAT ALPHA"80 OUTPUT 722;"CAL?";N,T90 ENTER 722; A$

Strona 83 - Appendix A Specifications

Chapter 5 Command Summary 73CALNUM?Description Calibration Number Query. Returns a decimal number indicating the number of times the multimeter has

Strona 84 - 1 / DC Voltage

74 Chapter 5 Command Summary CALSTRDescription Calibration String (remote only). Stores a string in the multimeter's nonvolatile calibration RAM

Strona 85 - 2 / Resistance

Chapter 5 Command Summary 75REV?Description Revision Query. Returns two numbers separated by a comma. The first number is the multimeter's outg

Strona 86 - 2 Accuracy

76 Chapter 5 Command Summary SCALDescription Service Calibration. Adjusts the AC sections of the instrument. Calculates the corrections to accurate

Strona 87 - 3 / DC Current

Chapter 5 Command Summary 77SECUREDescription Security Code. Allows the person responsible for adjustment to enter a security code to prevent accide

Strona 89 - Settling Characteristics

78 Chapter 5 Command Summary Disabling SecurityOUTPUT 722;"SECURE 3458,0" !DISABLES SECURITY FOR ADJUSTMENT AND AUTOCAL

Strona 90

Chapter 5 Command Summary 79TEMP?Description Temperature Query. Returns the multimeter's internal temperature in degrees Centigrade.Syntax TEMP

Strona 91

80 Chapter 5 Command Summary TESTDescription Causes the multimeter to perform a series of internal self-test.Syntax TESTRemarks • Always disconnect a

Strona 92

Appendix A Specifications 81Appendix A SpecificationsIntroductionThe 3458A accuracy is specified as a part per million (ppm) of the reading plus a

Strona 93 - 5 / AC Current

82 Appendix A Specifications 1 / DC VoltageDC VoltageAccuracy3 (ppm of Reading (ppm of Reading for Option 002) + ppm of Range)Transfer Accuracy/Linea

Strona 94 - 6 / Frequency/ Period

Appendix A Specifications 83Reading Rate (Auto-Zero Off) Temperature Coefficient (Auto-Zero off) For a stable environment ±1°C add the following ad

Strona 95 - 7 / Digitizing Specifications

84 Appendix A Specifications 2 Accuracy1 (ppm of Reading + ppm of Range)Two-Wire Ohms AccuracyFor Two-Wire Ohms ( OHM ) accuracy, add the following

Strona 96 - Voltage

Appendix A Specifications 853 / DC CurrentDC Current (DCI Function)Accuracy 3 (ppm Reading + ppm Range)Range Full ScaleMaximum ResolutionShunt Resis

Strona 97 - 8 / System Specifications

86 Appendix A Specifications 4 / AC VoltageGeneral InformationThe 3458A supports three techniques for measuring true rms AC voltage, each offering un

Strona 98 - 10 / Math Functions

Appendix A Specifications 87AC Accuracy (continued): 24 Hour to 2 Year (% of Reading + % of Range)Transfer AccuracyAC + DC Accuracy (ACDCV Function

Strona 99 - 11 / General Specifications

Chapter 1 3458A Calibration Introduction 7Chapter 1 3458A Calibration IntroductionIntroductionThis manual provides operation verification procedu

Strona 100

88 Appendix A Specifications High Frequency Temperature CoefficientFor outside Tcal ±5°C add the following error.(% of Reading)/°CAnalog Mode (ACV F

Strona 101 - 3458A (Product Note 3458A-3)

Appendix A Specifications 89Reading Rates 1Settling CharacteristicsFor first reading or range change error using default delays, add .01% of input s

Strona 102 - 10 kΩ resistance standard

90 Appendix A Specifications AC + DCV Accuracy (ACDCV Function)For ACDCV Accuracy apply the following additional error to the ACV accuracy. (% of Ra

Strona 103 - Offset Adjustments

Appendix A Specifications 915 / AC CurrentAC Current (ACI and ACDCI Functions)AC Accuracy 224 Hour to 2 Year (% Reading + % Range)AC + DC Accuracy (

Strona 104

92 Appendix A Specifications Settling CharacteristicsFor first reading or range change error using default delays, add .01% of input step additional

Strona 105

Appendix A Specifications 937 / Digitizing SpecificationsGeneral InformationThe 3458A supports three independent methods for signal digitizing. Eac

Strona 106 - Traceability

94 Appendix A Specifications Dynamic Performance100 mV, 1 V, 10 V Ranges; Aperture = 6 µsDirect and Sub-sampled Digitizing (DSDC, DSAC, SSDC and SSA

Strona 107 - Internal Reference Constants:

Appendix A Specifications 958 / System SpecificationsFunction-Range-MeasurementThe time required to program via GPIB a new measurement configuratio

Strona 108

96 Appendix A Specifications 9 / RatioType of Ratio 1Accuracy10 / Math FunctionsGeneral Math Function SpecificationsMath is executable as either a re

Strona 109 - *03458-90017*

Appendix A Specifications 9711 / General SpecificationsOperating EnvironmentTemperature Range: 0°C to 55°COperating Location: Indoor Use OnlyOperati