Functions and Features 2
34450A User’s Guide 63
In data log/external trigger mode, a trigger out signal is sent
whenever a measurement is log/triggered and updated on
the front panel.
In code compatibility mode, a trigger out signal is sent
whenever a measurement is updated on the front panel or a
measurement is taken by the user using command.The
diagram below shows the trigger out connector:
Figure 2-6 Trigger out connector
Manual Trigger
Manual triggering takes one reading (or the specified number
of readings in data logger) each time the multimeter receives
a pulse on the front- panel trigger button. The diagram below
shows the manual trigger operation:
The trigger annunciator turns on when the multimeter is
waiting for a manual trigger. The rear panel trigger out
connector provides a pulse after the completion of each
+
Tx
Rx
TRIG
IN
TRIG
OUT