*ESE <NRf>
The *ESE (Event Status Enable) <NRf> command sets the Standard Event
Status Enable Register. This register contains a mask value for the bits to
be enabled in the Standard Event Status Register. A 1 in the Enable
Register enables the corresponding bit in the Status Register, a 0 disables
the bit. The parameter value, when rounded to an integer and expressed
in base 2, represents the bit values of the Standard Event Status Enable
Register. Table 17- 90 shows the contents of this register.
Tab le 17 - 90 *ESE Mapping
Syntax
Bit Weight Meaning
0 1 Operation Complete
1 2 Request Control (not used)
2 4 Query Error
3 8 Device Dependent Error
4 16 Execution Error
5 32 Command Error
6 64 Not used
7128 Power On
*ESE
NRf
?
Space