SCPI Command Reference 6
N9340A User’s Guide 129
This command specifies whether the step size is
set automatically based on the span.
Center Frequency Step Size
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]
<freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
This command specifies the center frequency step
size.
Frequency Span
[:SENSe]:FREQuency:SPAN <freq>
[:SENSe]:FREQuency:SPAN?
This sets the frequency span.
Setting the span to 0 Hz puts the analyzer into
zero span.
Full Frequency Span
[:SENSe]:FREQuency:SPAN:FULL
This command sets the frequency span to full
scale.
*RST State: On
Key Access: [FREQ] > {CF Step}
*RST Value: 300 MHz
Available unit: Hz, kHz, MHz, GHz
Key Access: [FREQ] > {CF Step}
*RST Value: 3.0 GHz
Available unit: Hz, kHz, MHz, GHz
Key Access: [SPAN] > {Span}
*RST Value: 3.0 GHz
Key Access: [SPAN] > {Full}