5
N5230A/C Corrected System Performance with 3.5mm Connectors
1
Option 140 or 145 or 146 or 240 or 245 or 246
1
From 300 kHz to 10 MHz, performance is characterized as "typical". To generate these typical values, please download our free
Uncertainty Calculator from http://www.agilent.com/find/na_calculator
.
Note: For any S
ii
reflection measurement:
• S
jj
= 0.
For any S
ij
transmission measurement:
• S
ji
= S
ij
when S
ij
≤ 1
• S
ji
= 1/S
ij
when S
ij
≥ 1
• S
kk
= 0 for all k
Table 3. 85052B Calibration Kit
N5230A/C - Option 140 or 240 (Standard Test Set and Standard Power Range), Option 145 or 245
(Configurable Test Set and Extended Power Range), Option 146 or 246 (Configurable
Test Set and Extended Power Range with Internal Second Source)
Applies to the N5230A/C Option 140 or 145 or 146 or 240 or 245 or 246 analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port
cable set, and a full 4-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Description Specification (dB)
10 MHz to
500 MHz
500 MHz to
2 GHz
2 to
8 GHz
8 to
13.5 GHz
13.5 to
20 GHz
Directivity 48 48 44 44 44
Source Match 40 40 33 31 31
Load Match 48 48 44 44 44
Reflection Tracking
±0.003
+0.010/°C
±0.003
+0.010/°C
±0.003
+0.020/°C
±0.006
+0.030/°C
±0.006
+0.030/°C
Transmission Tracking
±0.017
+0.010/°C
±0.017
+0.010/°C
±0.062
+0.020/°C
±0.125
+0.030/°C
±0.125
+0.030/°C