Agilent Technologies L4400 Podręcznik Użytkownika Strona 264

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 281
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 263
252 L4400 User’s Guide
A L4451A and L4452A Calibration Procedures
Calibration Process
The following general procedure is the recommended method to complete
a full instrument calibration.
1 See “L4451A and L4452A Performance Test Considerations.”
2 Perform the verification tests to characterize the instrument (incoming
data).
3 Unsecure the instrument for calibration (“Calibration Security" on
page 250).
4 Perform the DAC adjustment procedures on either the L4451A or
L4452A.
5 Secure the instrument against unauthorized calibration.
6 Note the security code and calibration count in the instrument’s
maintenance records.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the procedure
has been initiated. You can abort a calibration at any time by turning off
the power. You can also abort the calibration by sending the device clear
message or the CALibration:ABORt command.
CAUTION
If you abort a calibration in progress by cycling power when the
instrument is attempting to write new calibration constants to EEPROM,
you may lose all calibration constants for the function. Typically, upon
re-applying power, the instrument will report error 705 Cal:Aborted.
You may also generate errors 740 through 746. If this occurs, you
should not use the instrument until a complete re-adjustment has been
performed.
Przeglądanie stron 263
1 2 ... 259 260 261 262 263 264 265 266 267 268 269 ... 280 281

Komentarze do niniejszej Instrukcji

Brak uwag