Agilent 34980A and L4445A
Configuring an RF/Microwave
Switch System
Application Note
Agilent Technologies
Contents
Introduction 2
Overview of 34980A RF/Microwave Switch Modules 3
Overview of 34945A/L4445A RF/Microwave Switch/Attenuator Driver 5
Key RF/Microwave Test System Specifications 7
Tips in building an RF/Microwave Test System 9
Key factors in selecting RF/Microwave Switches 10
Controlling Agilent Switches/Attenuators 11
34945EXT Distribution Board Configuration 13
Custom Switching and the 34945A/L4445A Switch Driver 24
Example Switch Driver Configurations 26
The Web Server Diagnostic Tool 28
Switch Executive vs. Agilent Switch Sequencer 29
Conclusion 31
34980A and
Plug-in Modules
L4400 Modules
Przeglądanie stron 0
1 2 3 4 5 6 ... 31 32

Podsumowanie treści

Strona 1 - Switch System

Agilent 34980A and L4445A Configuring an RF/Microwave Switch System Application NoteAgilent TechnologiesContentsIntroduction 2Overview of 34980A RF/Mi

Strona 2

6. Choose a switching sub-systemthat is easy to configure, verify,and provides the capacity toexpand your test system in thefuture. How you control ex

Strona 3

A test system with manyFailsafe-type switches requiresa significant amount of currentto maintain switch closures.Some RF /Microwave SwitchDrivers can

Strona 4

Position feedback is often astandard feature with Agilentswitches, because the construc-tion of the switch drive signalscan also feed back information

Strona 5

34945EXT Distribution Board Configuration Steps to select and configurethe 34945EXT associated distribution boards, relays,attenuators & cabling:I

Strona 6

14Table 1. Agilent Switch/Attenuator to Y115x Distribution BoardAgilent Recommended Switch Options Description Frequency 34945EXT Distribution Boa

Strona 7

15Table 1. Agilent Switch/Attenuator to Y115x Distribution Board (Continued)Agilent Recommended Switch Options Attenuator Frequency 34945EXT Distrib

Strona 8

16Table 2. Agilent Switch to Y1155 Distribution BoardAgilent Recommended Options Description Frequency 34945EXT Distribution Board Model Coil Indi

Strona 9

Tables 1-7 will describe which kits are appropriate for the various switches and attenuators.Agilent Y1157A Cable Kit• 9 to 10-pin cable kit for Y1150

Strona 10

Table 3. The Y1150A Distribution BoardY1150A Switch optionsCoil voltage (24 V) 124Position feedback 402Connector type (9-pin) 201Switch types: N1810U

Strona 11

Table 4. The Y1151A Distribution BoardY1151A Switch optionsCoil voltage (24 V) 024Position feedback Comes StandardConnector type (16-pin) 161Switch t

Strona 12

IntroductionLAN (Ethernet) is fast becomingthe basis for a new set ofinstrumentation called LXI, or LAN eXentions for Instru-mentation. LXI is a colle

Strona 13

Table 5. The Y1152A Distribution BoardY1152A Switch optionsCoil voltage (24 V) Comes StandardPosition feedback Comes StandardConnector type (16-pin)

Strona 14 - Option Option (style)

Table 6. The Y1153A Distribution BoardY1153A.1 Switch optionsCoil voltage (24 V) 024Position feedback Comes StandardConnector type (10-pin) Comes Sta

Strona 15

The Agilent 8494/5/6 attenua-tors come standard with aViking connector and a Vikingto bare wire cable. The Agilent8490x attenuators use a 10-pinsocket

Strona 16 - Voltage Option Option (style)

Table 7. The Y1154A Distribution BoardY1154A Switch optionsCoil voltage (24 V) Comes StandardPosition feedback Comes StandardConnector type (10-pin)

Strona 17

Custom Switching and the34945A/L4445A Switch DriverThe Y1150A through Y1154Aprovide very specific solutionsfor connecting to a wide rangeof Agilent sw

Strona 18

The 34945EXT only supportsCommon Positive coil drive, soyou must connect the commonconnection on the switch to the +VR lines on the Y1155A,and the swi

Strona 19

Example 34945A/L4445A Switch Driver ConfigurationsThe following two examplesshould give you a pretty goodidea of how to configure a typical RF/Microwa

Strona 20

Here is a possible Agilent34980A configuration with discussion points:• 34980A Switch/Measure Unit• (1) 34945A Switch/Attenuator Driver w/cable • (3)

Strona 21 - Y1153A.2 Switch options

The Web Server Diagnostic ToolThe 34980A/L4445A has abuilt-in Web Server that provides a very powerful configuration, diagnosis, andverification tool.

Strona 22 - Ribbon Cable:

When developing test systemprograms, you can open up aWeb Browser to the 34980A/L4445A, and you can observeswitches closing and configura-tions changi

Strona 23

However, you may need to rapidly deploy customRF/Microwave Switch solutionsfor your testing needs, so youneed to be in control of leadtimes and costs

Strona 24 - LED Drive

A switch executive is an external software package that provides the following features: • Switch management and report generation• Creating switch pa

Strona 25

ConclusionYou have been introduced to a number of 34980A/L4445Acapabilities related toRF/Microwave switching. You have gained a betterunderstanding of

Strona 26 - Tables 3 and 5

www.agilent.com/find/emailupdatesGet the latest information on the products and applications you select. www.agilent.com/find/agilentdirectQuickly cho

Strona 27 - Tables 1 and 2, the Agilent

Using RF and Microwave ModulesUsing 34980A mainframe modules to construct your high frequency test system has the following attributes:Benefits:• Conv

Strona 28 - The figure at right shows

Overview of 34945A/L4445RF/Microwave/Attenuator DriverThe 34945A/L4445A MicrowaveSwitch/Attenuator driver isdesigned to support virtuallyany RF/Microw

Strona 29

The previous figure showssome key aspects of the34945A/L4445A Switch Driver:• Each 34945A/L4445A module can supply 24 V to the first 34945EXT• Daisy c

Strona 30

A disadvantage of continuousdrive switches is they maychange performance due topower dissipation in the switch housing. This can leadto deviations in

Strona 31

Frequency rangeThe top frequency limits ofswitches are set by the size of the coaxial structure andconnectors. SMA connectorsoperate from DC to 20 GHz

Strona 32 - Agilent Technologies

Termination At high frequencies, all signalsmust be properly terminated.With improper termination(open end or mismatchedimpedance), signals can berefl