Agilent Technologies HP 8642A/B Dokumentacja Strona 153

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 453
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 152
Derive From FM Rate and Deviation
Introduction
The FM Rate and Deviation method is very accurate and is the only
fully automated method for determining the Discriminator Constant. This
calibration method requires the application of a frequency modulated
signal to the measurement configuration. The FM rate and deviation must
be determined. You will be required to define these values either in the
Calibration Process display or when the Connect Diagram appears.
The HP 3048A measures the spur level at the output of the discriminator
(Phase Detector) and calculates the Discriminator Constant.
Calibr Process: $N w/FM Discriminator 2-85
Przeglądanie stron 152
1 2 ... 148 149 150 151 152 153 154 155 156 157 158 ... 452 453

Komentarze do niniejszej Instrukcji

Brak uwag