Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 4- 55
Function Details
SMU Ranging Mode
SMU Ranging Mode
Agilent B1500 provides the following operation modes for the measurement range
and the output range. The ranging mode must be set for each channel.
“Auto Ranging”
“Limited Auto Ranging”
“Fixed Range”
“Compliance Range”
“Enhanced Auto Ranging for Current Measurement”
To Set Ranging Mode
The following points must be noted when setting the ranging mode.
For measurement
channels
If you do not specify the ranging mode, the auto ranging mode is set
automatically.
For the channels that perform Vf/Vm (voltage force and voltage measurement)
or If/Im (current force and current measurement), the measurement range is
always the same as the output range.
For measurement with pulse output, the measurement channel uses the
compliance range or the specified fixed range. The channel cannot perform the
auto ranging operation or the limited auto ranging operation.
For the SMU connected to the atto sense/switch unit (ASU), the 10 pA range is
the minimum for the auto ranging operation. To use the 1 pA range, select the 1
pA limited auto ranging mode or the 1 pA fixed range mode.
For output
channels
The auto ranging mode and the limited auto ranging mode are available for the
output channels.
When the voltage output range or the voltage compliance range is changed, the
channel forces 0 V at that moment.