Table 6. 85056D Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E8363B/C AND E8364B/C analyzers, 85056D (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
45 MHz to
2 GHz
2 to
20 GHz
20 to
40 GHz
40 to
50 GHz
Directivity 42 34 26 26
Source Match 40 30 24 23
Load Match 42 33 25 25
Reflection Tracking
±0.002
+0.02/°C
±0.029
+0.02/°C
±0.079
+0.02/°C
±0.075
+0.03/°C
Transmission Tracking
±0.022
+0.02/°C
±0.130
+0.02/°C
±0.365
+0.02/°C
±0.498
+0.03/°C