6
ProductSpecifications
Definitionsandconditions
Specification
Specificationsdescribetheperformanceparameterscoveredbytheproductwarrantyandare
validfrom20to35°Cunlessotherwisenoted.
Typical
Typicaldescribesadditionalproductperformanceinformationthatisnotcoveredbythe
productwarranty.Itisperformancebeyondspecificationsthat95percentoftheunitsexhibit
witha95percentconfidencelevel.Thisdata,showninitalics,doesnotincludemeasurement
uncertainty,andisvalidonlyatroomtemperature(approximately25°C)afteralignmentwithin
thestatedalignmenttimeandtemperaturelimits.
Nominal
Nominalvaluesindicateexpectedperformance,ordescribeproductperformancethatisuseful
intheapplicationoftheproduct,butarenotcoveredbytheproductwarranty.
Thetestsetwillmeetitsspecificationwhen:
• Thetestsetiswithinitscalibrationcycle
• Thetestsethasbeenstoredatanambienttemperaturewithintheallowedoperating
rangeforatleasttwohoursbeforebeingturnedon;ifithaspreviouslybeenstoredata
temperaturerangeinsidetheallowedstoragerange,butoutsidetheallowedoperating
range
• Thetestsethasbeenturnedonforatleast45minutes
• TheRF,IF,andSourceAlignments
1
havebeenrunwithintheprevious7days
• AnALLAlignment
1
hasbeenrun:
• Withintheprevious8hours
• Ifthetemperaturehaschangedmorethan5°Cfromtheprevious“ALL”alignment
1. For more information on using Alignments in a manufacturing environment, please see the EXM user documentation.