Specifications
Hardware Specifications
Specifications in Table A-1 are warranted. Specifications apply over an ambient temperature range of 0°
C to 40° C. When charging, specifications apply for charging voltages from 0.5 V to maximum, and
charge currents from minimum to maximum. When discharging, specifications apply for discharging
voltages from 1.5 V to maximum and discharging currents from minimum to maximum. Accuracy
specifications apply over the entire range of ac line and power bus conditions (line regulation), and
charging/discharging levels (load regulation). Specifications are subject to change without notice.
Table A-1. Agilent E4370A/E4374A/E4375A MCCD Specifications
Parameter Condition E4374A Value E4375A Value
Maximum Programmable
Output Voltage:
charging, measured at cell 5 V 5 V
Maximum Compliance Voltage:
(cell voltage + fixture/wiring
voltage drops)
charging, measured at
MCCD connector
5.5 V 6.0 V
Maximum Programmable
Output Current:
charging or discharging,
per channel
2 A 3 A
Maximum Output Current:
1
charging or discharging,
constant voltage
0 A 0 A
Maximum Output Leakage
Current:
disabled, per channel, with an
external voltage of -5V to +5 V
± 25 µA ± 25 µA
Maximum Power:
charging, per channel
discharging, per channel
11 W
9 W
18 W
13.5 W
Maximum Input Voltage:
discharging 4.5 V 4.5 V
Voltage Programming Accuracy:
measured at sense connector
input with remote sensing
± 1 mV ± 1 mV
Voltage Readback Accuracy:
measured at sense connector
input with remote sensing
± 1 mV ± 1 mV
Current Programming
Accuracy:
% of reading + offset 1A
> 1A to 2A
> 2A
± (0.05% + 1mA)
± (0.10% + 1mA)
N/A
± (0.05% + 1.5mA)
± (0.10% + 1.5mA)
± (0.15% + 1.5mA)
Current Readback Accuracy:
% of reading + offset 1A
> 1A to 2A
> 2A
± (0.05% + 1mA)
± (0.10% + 1mA)
N/A
± (0.05% + 1.5mA)
± (0.10% + 1.5mA)
± (0.15% + 1.5mA)
ac Resistance Measurement
Accuracy:
% of reading + offset
± (1% + 1m) ± (1% + 1m)
dc Resistance Measurement
Accuracy:
% of reading + offset
± (1% + 1m) ± (1% + 1m)
1
There is a minimum programmable current limit when operating in constant current (CC) mode. In CC mode, the output cannot
be set to run below the minimum programmable constant current limit specified in Table A-2. In constant voltage (CV)mode,
charging or discharging, the MCCD will regulate current down to 0A.