Agilent Technologies B1500A Podręcznik Użytkownika Strona 152

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 202
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 151
4- 12 Agilent B1500A User’s Guide, Edition 13
Specifications
Hardware
HPSMU
HPSMU: High Power Source/Monitor Unit
Voltage range, resolution, and accuracy
See Table 4-5 for high resolution ADC and Table 4-7 for high speed ADC.
Current range, resolution, and accuracy
See Table 4-6 for high resolution ADC and Table 4-8 for high speed ADC.
Specified measurement resolution is limited by fundamental noise limits.
Power consumption
Voltage source mode:
Current source mode:
Voltage range (V)
Power consumption
a
(W)
a. Ic is the current compliance setting.
2 Ic 20
20 Ic 20
40 Ic 40
100 Ic 100
200 Ic 200
Voltage compliance
a
(V)
a. Vc is the voltage compliance setting.
Power consumption
b
(W)
b. Io is the output current.
Vc  20 Io 20
20 < Vc 40 Io 40
40 < Vc 100 Io 100
100 < Vc 200 Io 200
Przeglądanie stron 151
1 2 ... 147 148 149 150 151 152 153 154 155 156 157 ... 201 202

Komentarze do niniejszej Instrukcji

Brak uwag