1-2 85032B/E
General Information
Calibration Kit Overview
Calibration Kit Overview
The Agilent 85032B and 85032E type-N calibration kits are used to calibrate Agilent
network analyzers up to 6 GHz for measurements of components with 50Ω type-N
connectors.
Kit Contents
Use the Contents List in the shipping container to verify the completeness of your
shipment. Although this list is the most accurate, you can also use the illustrations in
Chapter 7 to verify the items in your shipment. If your shipment is not complete, contact
Agilent Technologies - refer to “Contacting Agilent” on page 5-4.
Broadband Loads
The broadband loads are instrument-grade, 50Ω terminations that have been optimized for
performance up to 6 GHz. The rugged internal structure provides for highly repeatable
connections. A distributed resistive element on sapphire provides excellent stability and
return loss.
Opens and Shorts
The opens and shorts are built from parts that are machined to the current state-of the-art
precision machining.
The short’s inner conductors have a one-piece construction, common with the shorting
plane. This construction provides for extremely repeatable connections.
The female open has a separate-piece inner conductor that is made from a
low-dielectric-constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very
tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The
lengths of these offsets are designed so that the difference in phase of their reflection
coefficients is approximately 180 degrees at all frequencies.
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide
good broadband performance. The adapters utilize a dual-beaded connector structure to
ensure stable, repeatable connections. The beads are designed to minimize return loss and
are separated far enough so that interaction between the beads is minimized.
The adapters are designed so that their nominal electrical lengths are the same, which
allows them to be used in calibration procedures for non-insertable devices.
Calibration Definitions
The calibration kit must be selected and the calibration definitions for the devices in the
kit installed in the network analyzer prior to performing a calibration. Refer to your
network analyzer users guide for instructions on selecting the calibration kit and
performing a calibration.