Agilent Technologies 42A Podręcznik Użytkownika Strona 123

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 194
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 122
Te st R e c o rd Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide, Sixth Edition 123
Agilent 81680A Performance Test
Page 11 of 13
Model Agilent 81680A Tunable Laser Report No. ________ Date_______
Power Stability
1
Power Stability = Maximum Deviation – Minimum Deviation
Signal-to-Source Spontaneous Emission - 81680A
Output 2, High Power
Measurement Uncertainty: ±0.20 dB
Low SSE Output 1 High Power Output 2
Att = 0 dB
Maximum Deviation dB dB
Minimum Deviation dB dB
Power Stability
1
dB dB
Specification 0.02 dBpp 0.02 dBpp
Measurement Uncertainty ±0.005 dB ±0.005 dB
Wavelength Output Power Results
Maximum
Specification
1460 nm -3.00 dBm dB 35 dB
1470 nm -3.00 dBm dB 35 dB
1480 nm +1.00 dBm dB 40 dB
1490 nm +1.00 dBm dB 40 dB
1500 nm +1.00 dBm dB 40 dB
1510 nm +1.00 dBm dB 40 dB
1520 nm +5.00 dBm dB 45 dB
1530 nm +5.00 dBm dB 45 dB
1540 nm +5.00 dBm dB 45 dB
1550 nm +5.00 dBm dB 45 dB
1560 nm +5.00 dBm dB 45 dB
1570 nm +5.00 dBm dB 45 dB
1580 nm +1.00 dBm dB 40 dB
Przeglądanie stron 122