Seconds Per Division Field (Timing only)
When the analyzer Type field in the Configuration menu is set to Timing, the
analyzer uses its own internal clock. In this mode, the X-axis of the
waveform display is measured in seconds per division.
You use the sec/Div field to select the seconds per division resolution of the
X-axis. The range of the sec/Div field is 1 ns/Div to 1.0 ks/Div. You set the
sec/Div field by either selecting the sec/Div field and rotating the knob, or by
highlighting the sec/Div field and entering a value using the front-panel
keypad.
When using the knob to set the sec/Div, the value will change in a 1-2-5
sequence. By adjusting the sec/Div, you can zoom in to view a desired part of
the display.
Seconds Per Division Field
Seconds per
division field
The Waveform Menu
Seconds Per Division Field (Timing only)