Chapter 5 81
Spectrum Analyzer
Measuring Multiple Signals
Spectrum Analyzer
Press FREQ Channel, CF Step, 10, MHz.
Step 7. Activate the marker delta function:
Press
Marker, Delta.
Step 8. Increase the center frequency by 10 MHz:
Press
FREQ Channel, Center Frequency, , Peak Search.
The delta marker (1) appears on the peak of the 40 MHz harmonic.
The delta marker annotation displays the amplitude and frequency
difference between the 30 and 40 MHz signal peaks. Refer to
Figure 5-6.
Figure 5-6 Delta Marker with Reference Signal Off-Screen
Step 9. Turn the markers off:
Press
Marker, Off.