7-6 85330A Multiple Channel Controller
Remote Programming Command Reference
Standard Command Reference
ROUTe:CONTrol
Controls the TTL state of the rear panel AUX 1 connector.
Default Default is a TTL LOW.
Syntax OUT:CONT <data>;
Example ROUT:CONT 0;
Sets AUX 1 to TTL LOW
ROUT:CONT 1;
Sets AUX 1 to TTL HIGH
NOTE The state of AUX 1 must be set before entering run-time mode.
ROUTe:DELay
Adds a specific amount of delay between a switch address entry being
asserted and the issuance of a measurement trigger. Delay is entered in
micro-seconds.
Default The default is 160 micro-seconds.
Syntax ROUT:DEL <micro-seconds>;
Example ROUT:DEL 200;
Set the trigger delay (settling time) to 200 micro-seconds.
NOTE Most solid state switches are capable of changing states within a few
micro-seconds.