98 Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
Programming Examples
GPIB Programming Interface Examples
Using 8757D Pass-Thru Commands (PSG with Option 007 Only)
Pass- thru commands enable you to send operating instructions to a PSG that is connected to a
8757D scalar analyzer system. This section provides setup information and an example program for
using pass- thru commands in a ramp sweep system.
Equipment Setup
To send pass- thru commands, set up the equipment as shown in Figure 3- 1. Notice that the GPIB
cable from the computer is connected to the GPIB interface bus of the 8757D. The GPIB cable from
the PSG is connected to the system interface bus of the 8757D.
Figure 3-1